  • IRW-1000 Infant Radiant Warmer
IRW-1000 Infant Radiant Warmer

IRW-1000 Infant Radiant Warmer

Standard composition
1. Servo Control Mode: air temperature
2. Instant Parameter Display: set temp, skin temp, heat power rate;
3. Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
4. Tilting Bed with Rotary Radiant Head
5. X-ray Cassette Tray under infant bed, examination light of high illuminance
6. Use anti-blast quartz infrared tube as radiant heating source.


Technical specification:
Power :

Model IRW-1000A-1E AC 110 / 120 V, 50Hz / 60Hz;
Model IRW-1000A-2E AC 220 / 240V, 50Hz / 60Hz;
Input Power ≤1000VA;
Temperature control range and relative specification :

Control mode Auto mode;
Temperature control range 25°C~37°C;
Skin temperature sensor measure range 20°C~42°C;
Deviation between sensor measure temperature and control temperature ≤0.5°C;
Skin Sensor Precision ≤0.3°C;
Bed surface temperature Uniformity ≤1°C;
Temperature Rise Time ≤45min;
Bed tiltable continues adjustable in ±10°;
Alarm :

Sensor failure alarm: Alarm with audible and visual when sensor open circuit, short circuit or disconnect, cut off power;
Over temperature alarm: Alarm with audible and visual when displayed temperature was approach 38°C;
Deviation alarm :After temperature equilibrium, alarm with audible and visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C;
Power failure alarm :Alarm with audible and visual when power failure;
Environment Temperature (It is a suggestion that please do not use outside of the allow environment.):

Operate temperature range 18°C~30°C;
Store temperature range -20°C~+55°C;
Environment Humidity :

Operate humidity range 30%RH~75%RH;
Store humidity range ≤93%RH;
Atmospheric pressure :

Transport and Store atmospheric pressure range 500hPa~1060hPa;
Operate atmospheric pressure range 700hPa~1060hPa;
Stream velocity :

Ambit environment stream velocity ≤0.3m/s。