Alarm :
Power Alarm :Power Failure: Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle;
Fan Alarm : Fan Failure: Activates if Fan motor malfunctions;Sensor Alarm :Air Sensor Failure: Air Sensor malfunction, refer unit to service;
Deviation Alarm :High Air Temperature: Activated if displayed Air Temperature is 15°C above set point;Low Air Temperature: Activated if displayed Air Temperature is 25°C below set point;
Over Temperature Alarm :Over Air Temperature: Activated when displayed Air Temperature was approach 38°C for set temperature ≤37°C, or approach 40°C for set temperature >37°C in Air Mode;Temperature Control Range and Correlative Specifications :
Air Temperature :20°C~37°C,for set temperature ≤37°C, 37°C~39°C, for set temperature >37°C;
Temperature Rise Time :≤30min;Environment Requirement :
Environment Temperature: Operation Temperature Range +20°C~+30°C;Store Temperature Range -40°C~+55°C;
Environment Humidity: Operation Humidity Range 30%~75%;Store Humidity Range 10%~100%;Other Specifications :
Noise Level Within Hood Environment :≤55dB(Environment noise below 45dB);
Carbon Dioxide (CO2) Level within the Hood Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress;Function of "SILENCE" Key :
Make the alarm to be silence for 10 minutes by press this key :Deviation Alarm;Sensor Alarm;Fan Alarm;
Factory Default Settings :
Temperature Control Mode Air Temperature Mode;Air Set Temperature: 32°C;Temperature Units:°C